• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INSTITUT STRAUMANN AG TE Ø 4.8MM WN, SLA® 12MM, TIZR, NTP; ENDOSSEOUS DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INSTITUT STRAUMANN AG TE Ø 4.8MM WN, SLA® 12MM, TIZR, NTP; ENDOSSEOUS DENTAL IMPLANT Back to Search Results
Catalog Number 043.661S
Medical Device Problem Code Difficult to Insert (1316)
Health Effect - Clinical Code Fibrosis (3167)
Date of Event 01/09/2026
Type of Reportable Event Serious Injury
Event or Problem Description
The clinician reports that the day the implant was placed in ada 30, failure occurred upon insertion.Details of surgery: primary stability not achieved and implant surface not completely covered with bone.Patient presented with bone type iii and inadequate bone quality/quantity.The device was forwarded to the manufacturer.At the event the patient experienced: mobility.No further patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TE Ø 4.8MM WN, SLA® 12MM, TIZR, NTP
Common Device Name
ENDOSSEOUS DENTAL IMPLANT
Manufacturer (Section D)
INSTITUT STRAUMANN AG
peter merian-weg 12
basel 04002
SZ  04002
MDR Report Key24485529
Report Number0001222315-2026-000596
Device Sequence Number14728680
Product Code DZE
UDI-Device Identifier07630031719270
UDI-Public07630031719270
Combination Product (Y/N)N
Initial Reporter StateCT
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Importer
Initial Reporter Occupation Dentist
Type of Report Initial
Report Date (Section B) 03/02/2026
Report Date (Section F) 02/06/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number043.661S
Device Lot NumberMLC87
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/06/2026
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Distributor Facility Aware Date03/02/2026
Device Age40 MO
Event Location Other
Date Report to Manufacturer02/06/2026
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date03/02/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age49 YR
Patient SexMale
-
-