| Catalog Number |
BN7TCDF4L |
| Medical Device Problem Code |
Patient Device Interaction Problem (4001)
|
| Health Effect - Clinical Code |
Heart Block (4444)
|
| Date of Event |
02/13/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
A patient underwent an atrioventricular nodal reentrant tachycardia (avnrt) ablation procedure with ez steer nav catheter and the patient experienced av heart block treated with pacemaker implantation.The report indicated that during the avnrt procedure, the patient suffered complete av block and needed a pacemaker.Shortly before block only slow junctional beats were seen which led the physician to believe we were on the right track to ablate the accessory pathway.From the moment of the first blocked beat to ablation stop 5-10 seconds passed.During ablation the temperature did not exceed 46°c and no impedance rise was observed.The surgery was not delayed.The procedure was not successfully completed.Multiple attempts have been made to gain clarification and additional information about this event with no response.Should any new information be obtained it will be assessed and processed accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster inc., or its employees that the report constitutes an admission that the product, biosense webster inc., or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's ref.# pc-(b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
On 11-mar-2026, the bwi product analysis lab received the complaint device for evaluation.The product analysis has begun but is not completed at this time.When the investigational analysis has been completed, a supplemental 3500a report will be submitted.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's ref.# (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
A patient underwent an atrioventricular nodal reentrant tachycardia (avnrt) ablation procedure with ez steer nav catheter and the patient experienced av heart block treated with pacemaker implantation.Device evaluation details: the product was returned to johnson & johnson medtech (j&j) for evaluation.A visual inspection and revision of all features were performed following johnson & johnson medtech procedures.Visual analysis revealed no damage or anomalies on the device.The device features were reviewed.No magnetic, force, irrigation, deflection, temperature or electrical issues were found during the product investigation.A manufacturing record evaluation was performed for the finished device batch number, and no internal actions were identified.The root cause of the adverse event remains unknown.In addition, no device malfunction was reported, there may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the analysis.The instruction for use (ifu) states that careful catheter manipulation must be performed to avoid cardiac damage, perforation, or tamponade.As part of the johnson & johnson medtech quality process, all devices are manufactured, inspected, and released to approved specifications.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's ref.# (b)(4).
|
| |
|