• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND - 3015516266 EMSYS TRL LNR N 52X36; HIP INSTRUMENT - ACETABULAR TRIAL LINER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 3015516266 EMSYS TRL LNR N 52X36; HIP INSTRUMENT - ACETABULAR TRIAL LINER Back to Search Results
Catalog Number 482252036
Medical Device Problem Codes Break (1069); Crack (1135)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/17/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that emphasys trial liners have chips and cracks.The issue was found during use.There was no reported patient or user harm, but a surgical delay occurred during the procedure.
 
Additional Manufacturer Narrative
Product complaint: (b)(4).Investigation summary
=
> according to the information received, it was reported that the chips and cracks in 2 emphasis trial liners.The product was not returned to depuy synthes; however, photos were provided for review.See attachment ([(b)(4).Device photo ad (b)(6) 2026]).The photo investigation revealed that the emsys trl lnr n 52x36 is chipped at the rim, no crack could be observed.Additionally, scratches were found on the convex section.The condition found can be attributed to a combination of heavy use and exposure to unintended forces during repeated attempts at assembly and disassembly with the mating device and other tools and hard edges coming in contact with the device.As the device was not returned, and as-received condition could not be assessed, a dimensional inspection and document/specification review were not completed.The overall complaint was confirmed as the observed condition of the emsys trl lnr n 52x36 would have contributed to the complained device issue.Based on the investigation findings, the potential cause is traced to unintended use error, and it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot
=
> the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EMSYS TRL LNR N 52X36
Common Device Name
HIP INSTRUMENT - ACETABULAR TRIAL LINER
Manufacturer (Section D)
DEPUY IRELAND - 3015516266
loughbeg ringaskiddy co.
cork
EI 
Manufacturer Contact
elba bello
700 orthpaedic dr.
warsaw, IN 46581
9083863534
MDR Report Key24490182
Report Number1818910-2026-03931
Device Sequence Number16123683
Product Code HWT
UDI-Device Identifier10603295510413
UDI-Public(01)10603295510413(10)891531
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number482252036
Device Lot Number891531
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/17/2026
Supplement Date Received by Manufacturer04/28/2026
Initial Report FDA Received Date03/02/2026
Supplement Report FDA Received Date04/30/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/07/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-