| Model Number |
DS560TBT |
| Medical Device Problem Codes |
Degraded (1153); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/08/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This notification was initiated to evaluate a returned device in relation to the sound abatement foam recall.During the review, it was identified that an incorrect blfm (blower foam) code had been entered in trackwise (tw).A review of the source documentation in sap confirmed that the blfm code was applied in error.The associated complaint information was further assessed and confirmed that no death or serious injury was reported in connection with this event.Additionally, there is no evidence of a malfunction that would be likely to lead to death or serious injury if it were to recur.Based on the available information and in accordance with applicable regulatory reporting requirements, this event does not meet the criteria for reportability to any regulatory authority.Therefore, no regulatory reporting is required at this time.
|
| |
|
Event or Problem Description
|
|
A remstarauto ht hum device was returned to the third-party service center in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.There was no initial alleged complaint found.During evaluation no foam particles were noted in the airpath, yet foam degradation was noted.The devices sound abatement foam was replaced to address the issue.Additionally, the device was scrapped by the service center after the evaluation was completed.
|
| |
|
Additional Manufacturer Narrative
|
|
The capa process identified complaints not being created for quality issues found during third party servicing.Based on a review of records impacted by the identified issue, it was determined that this record needed to be corrected and is now being submitted accordingly.
|
| |
|
Search Alerts/Recalls
|