• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. REMSTARAUTO HT HUM; VENTILATOR, NON-CONTINUOUS (RESPIRATOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. REMSTARAUTO HT HUM; VENTILATOR, NON-CONTINUOUS (RESPIRATOR) Back to Search Results
Model Number DS560TBT
Medical Device Problem Codes Degraded (1153); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/08/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This notification was initiated to evaluate a returned device in relation to the sound abatement foam recall.During the review, it was identified that an incorrect blfm (blower foam) code had been entered in trackwise (tw).A review of the source documentation in sap confirmed that the blfm code was applied in error.The associated complaint information was further assessed and confirmed that no death or serious injury was reported in connection with this event.Additionally, there is no evidence of a malfunction that would be likely to lead to death or serious injury if it were to recur.Based on the available information and in accordance with applicable regulatory reporting requirements, this event does not meet the criteria for reportability to any regulatory authority.Therefore, no regulatory reporting is required at this time.
 
Event or Problem Description
A remstarauto ht hum device was returned to the third-party service center in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.There was no initial alleged complaint found.During evaluation no foam particles were noted in the airpath, yet foam degradation was noted.The devices sound abatement foam was replaced to address the issue.Additionally, the device was scrapped by the service center after the evaluation was completed.
 
Additional Manufacturer Narrative
The capa process identified complaints not being created for quality issues found during third party servicing.Based on a review of records impacted by the identified issue, it was determined that this record needed to be corrected and is now being submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REMSTARAUTO HT HUM
Common Device Name
VENTILATOR, NON-CONTINUOUS (RESPIRATOR)
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 
7247330200
MDR Report Key24490324
Report Number2518422-2026-007261
Device Sequence Number20078595
Product Code BZD
UDI-Device Identifier00606959030510
UDI-Public(01)00606959030510
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K131982
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Y
Reporter Type Manufacturer
Report Source Other,Distributor
Initial Reporter Occupation 505
Remedial Action Recall
Type of Report Initial,Followup
Report Date (Section B) 05/08/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberDS560TBT
Device Catalogue NumberDS560TBT
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/08/2025
Supplement Date Received by Manufacturer05/08/2026
Initial Report FDA Received Date03/02/2026
Supplement Report FDA Received Date05/08/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured12/23/2015
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-1974-2021
Patient Sequence Number1
Patient SexUnknown
-
-