The manufacturer was contacted in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.A remstar pro cflex + device was returned to manufacturer service center with no initial alleged complaint.The device was visually inspected and found evidence visible foam particles/degradation.The device was scrapped after the evaluation was completed.
|
|
In previously submitted report, product brand name, device serviced by 3rdp and device available for eval in box d and occupation in box e, report source in box g, reason device not eval, health impact grid and evaluation method code grid in box h were incorrect.In this report, product brand name, device serviced by 3rdp and device available for eval in box d and occupation in box e, report source in box g, reason device not eval, health impact grid and evaluation method code grid in box h has been updated/corrected.The affected products with field complaints alleging pe-pur foam degradation have undergone the establishment of complaint codes specifically identifying foam degradation, for system one.A field correction action(2021-06-a) was initiated, and medical device recall letters were issued.A field action was initiated to replace pe-pur foam in affected products; degradation was added to the respective dfmeas and the risk management file ¿ 2203077 v27 (systemone) was updated to include hazards associated with foam degradation based on complaints analyzed through various health hazard evaluations (hhes).Risk tags ¿ degradation01, toxic02 and particulate01 reflect the safety risk of design containing the pe-pur foam.These complaints were monitored and trended in accordance with the complaint trending procedures.If the complaints were determined to meet the criteria for escalation, they were escalated for risk assessment through the post market clinical escalation process.As of the date of submission for this report, there is no indication that complaints for the failure in question have led to unacceptable levels of severity or probabilities of harm, and therefore no further action will be pursued.
|