This supplemental report is being submitted to provide the results of the final investigation.Updated field: d8, d9, h2, h3, h6, h11 the device was returned to olympus for inspection, and the reported failure was confirmed.In addition, the following reportable malfunctions were identified during the device evaluation: intermit or flicker image a root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: the complaint is confirmed, as the device exhibited evidence of intermittent/flickering image, which is consistent with the reported event and the most probable cause traced due to a component failure.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|