| Model Number |
15 |
| Medical Device Problem Code |
Defibrillation/Stimulation Problem (1573)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/25/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer contacted stryker to report a non-critical issue with their device.Upon evaluation of the customer¿s device, stryker observed that the device had logged an event code in the memory which is related to a potential issue of the device to deliver energy.There was no patient involvement reported with the event.
|
| |
|
Additional Manufacturer Narrative
|
|
Stryker evaluated the customer's device and verified the reported issue.The event code was cleared from the memory of the device and thereafter did not return.After observing proper device operation through functional and performance testing the device was returned to the customer for use.The cause of the reported issue could not be conclusively determined.
|
| |
|
Search Alerts/Recalls
|