• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WAVELIGHT GMBH WAVELIGHT FS200 FEMTOSECOND LASER; POWERED LASER SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WAVELIGHT GMBH WAVELIGHT FS200 FEMTOSECOND LASER; POWERED LASER SURGICAL INSTRUMENT Back to Search Results
Catalog Number 8065990941
Medical Device Problem Code Unexpected Therapeutic Results (1631)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/26/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H.3., h.6.: investigation, including root cause analysis, is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A healthcare professional reported that during creation of the flap, the corneal flap was thin near the side cut at the ten o¿clock position, adhesion at flap edge was difficult to separate the flap in the unknown eye of a patient, during refractive surgery.The symptoms were resolved and the patient were now recovered.
 
Additional Manufacturer Narrative
Additional information provided in d.9., h.3., h.6., and h.11.A review of the device history record (dhr) traceable to the reported serial number indicates that the product was processed and released according to the product¿s acceptance criteria.A review of the technical service onsite history showed no abnormalities that could have contributed to this event: system was successfully verified after the surgery date.Most recent onsite visit from field service engineer performed and signed service installation record.System meets specification as per service installation record.During onsite visit the field service engineer replaced application interface and beam expander.Field service engineer performed system verification as per service installation record.The review of logfile for the day of treatment shows all laser system functions were within specifications.The treatment of the patient's left eye was finished successfully without any technical issue.The thickness of the flap was thinner than the recommended flap thickness.No technical root cause could be identified during logfile review.The system was working within specification.The replaced part is expected to be returned to manufacturer for evaluation but has not been returned.No device related issues were identified based on the provided data.The investigation showed patient and/or handling related factors, that may contribute to an outcome similar to the reported event.Therefore, the case is coded as "inconclusive - other".The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WAVELIGHT FS200 FEMTOSECOND LASER
Common Device Name
POWERED LASER SURGICAL INSTRUMENT
Manufacturer (Section D)
WAVELIGHT GMBH
am wolfsmantel 5
erlangen 91058
GM  91058
Manufacturer (Section G)
WAVELIGHT GMBH
am wolfsmantel 5
erlangen 91058
GM   91058
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key24496157
Report Number3003288808-2026-00129
Device Sequence Number18795936
Product Code GEX
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K101006
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/16/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number8065990941
Device Lot Number16LAEK
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/08/2026
Supplement Date Received by Manufacturer06/26/2026
Initial Report FDA Received Date03/03/2026
Supplement Report FDA Received Date07/16/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/06/2024
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-