| Model Number |
H749518100 |
| Medical Device Problem Code |
Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/27/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that a catheter issue occurred.The 80% stenosed, 28 x 3.0 mm target lesion was located in the moderately tortuous and moderately calcified left anterior descending (lad).An opticross catheter was selected for intravascular ultrasound examination (ivus) of the target lesion.During the procedure, the tip became twisted.The procedure was completed with another of same device.The patient condition following the procedure was stable.
|
| |
|
Additional Manufacturer Narrative
|
|
E1: initial reporter phone number - (b)(6).
|
| |
|
Event or Problem Description
|
|
It was reported that a catheter issue occurred.The 80% stenosed, 28 x 3.0mm target lesion was located in the moderately tortuous and moderately calcified left anterior descending (lad).An opticross catheter was selected for intravascular ultrasound examination (ivus) of the target lesion.During the procedure, the tip became twisted.The procedure was completed with another of same device.The patient condition following the procedure was stable.
|
| |
|
Additional Manufacturer Narrative
|
|
E1: initial reporter phone number - (b)(6).Investigation results: device analysis findings: the device was not returned for analysis; therefore, a technical analysis could not be performed.Device history record (dhr) review: it was confirmed that this device met manufacturing specifications prior to distribution and there were no manufacturing deviations which could have contributed to the reported event.Investigation conclusion: this investigation has been assigned an investigation conclusion code of cause not established following a review of the reported event details, available information, and product record review.In this case the device was not returned for product analysis therefore limiting the identification of a most probable cause of the reported event.Improper handling, device manipulation or not following the correct instructions during the procedure, alongside the anatomical conditions of the patient, could be contributing factors to the reported issue.However, there is no additional detail pertinent to the event.
|
| |
|