| Model Number |
693558 |
| Medical Device Problem Code |
Over-Sensing (1438)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/12/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Continuation of d10: 407645 -lead - implant date: (b)(6) 2010 ; dtpa2q1 -crt-d - implant date: (b)(6) 2025.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that the implantable cardioverter defibrillator (icd) delivered inappropriate anti-tachycardia pacing (atp) and shocked for atrial tachycardia/atrial fibrillation (at/af) with rapid ventricular response which the device detected and treated as ventricular tachycardia (vt).The right ventricular (rv) lead had exhibited oversensing resulting in far field r-wave (ffrw) oversensing.The device and lead remains in use.No further patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Product event summary: the lead was not returned for analysis, however, performance data collected from the device was received and analyzed.Analysis of the device memory indicated t-wave oversensing.Analysis of the device memory indicated oversensing associated with the right ventricular lead.Correction: b5.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that the implantable cardioverter defibrillator (icd) delivered inappropriate anti-tachycardia pacing (atp) and shocked for atrial tachycardia/atrial fibrillation (at/af) with rapid ventricular response which the device detected and treated as ventricular tachycardia (vt).The right ventricular (rv) lead had exhibited oversensing resulting in t-wave oversensing (twos).The device and lead remains in use.No further patient complications have been reported as a result of this event.It was further reported via follow-up with the clinic that the patient had been hospitalized and their medications were adjusted.
|
| |
|