Model Number 693558
Medical Device Problem Code Over-Sensing (1438)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/12/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Continuation of d10: 407645 -lead - implant date: (b)(6) 2010 ; dtpa2q1 -crt-d - implant date: (b)(6) 2025.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
It was reported that the implantable cardioverter defibrillator (icd) delivered inappropriate anti-tachycardia pacing (atp) and shocked for atrial tachycardia/atrial fibrillation (at/af) with rapid ventricular response which the device detected and treated as ventricular tachycardia (vt).The right ventricular (rv) lead had exhibited oversensing resulting in far field r-wave (ffrw) oversensing.The device and lead remains in use.No further patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Product event summary: the lead was not returned for analysis, however, performance data collected from the device was received and analyzed.Analysis of the device memory indicated t-wave oversensing.Analysis of the device memory indicated oversensing associated with the right ventricular lead.Correction: b5.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
It was reported that the implantable cardioverter defibrillator (icd) delivered inappropriate anti-tachycardia pacing (atp) and shocked for atrial tachycardia/atrial fibrillation (at/af) with rapid ventricular response which the device detected and treated as ventricular tachycardia (vt).The right ventricular (rv) lead had exhibited oversensing resulting in t-wave oversensing (twos).The device and lead remains in use.No further patient complications have been reported as a result of this event.It was further reported via follow-up with the clinic that the patient had been hospitalized and their medications were adjusted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPRINT QUATTRO SECURE S MRI SURESCAN
Common Device Name
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT)
Manufacturer (Section D)
MEDTRONIC PUERTO RICO VILLALBA
road 149 km 56.3
villalba PR 00766
Manufacturer (Section G)
MEDTRONIC PUERTO RICO VILLALBA
road 149 km 56.3
villalba PR 00766
Manufacturer Contact
gerard torenvliet
8200 coral sea st ne
mounds view, MN 55112-4391
7635269277
MDR Report Key24497265
Report Number2649622-2026-06132
Device Sequence Number14728639
Product Code LWS
UDI-Device Identifier00613994446343
UDI-Public(01)00613994446343(17)121011
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
P920015
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2010
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/19/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date10/11/2012
Device Model Number693558
Device Catalogue Number693558
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/12/2026
Supplement Date Received by Manufacturer03/04/2026
Initial Report FDA Received Date03/03/2026
Supplement Report FDA Received Date03/19/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/01/2010
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
429878 LEAD
Outcome Attributed to Adverse Event Life Threatening; Hospitalization; Required Intervention;
Patient Age42 YR
Patient SexMale
-
-