| Model Number |
9-VSD-MUSC-012 |
| Medical Device Problem Code |
Difficult to Fold, Unfold or Collapse (1254)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/06/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
|
| |
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2026, a 12mm amplatzer muscular vsd occluder was chosen for a procedure using a 7f amplatzer torqvue delivery system.During procedure, the device do not take proper shape.The deformed device was never released from the delivery cable while inside the patient.There was no report of any interaction with cardiac structures during deployment and no report of any angulation/kink noticed with the delivery system.The procedure got aborted.The patient remained hemodynamically stable throughout the procedure.There was no clinically significant delay in procedure and no adverse patient effects.The patient was reported stable.
|
| |
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2026, a 12mm amplatzer muscular vsd occluder was chosen for a procedure using a 7f amplatzer torqvue delivery system.During procedure, the device do not take proper shape.The deformed device was never released from the delivery cable while inside the patient.There was no report of any interaction with cardiac structures during deployment and no report of any angulation/kink noticed with the delivery system.The procedure got aborted.The patient remained hemodynamically stable throughout the procedure.There was no clinically significant delay in procedure and no adverse patient effects.The patient was reported stable.
|
| |
|
Additional Manufacturer Narrative
|
|
An event of device deformity could not be confirmed.The device was returned to abbott for investigation and the device met functional specifications when analyzed under non-physiological conditions.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed, and the product met all specifications.Additionally, the lot was reviewed for similar complaints, and there is no indication of a lot-specific product issue.The cause of the reported device deformity could not be conclusively determined.There is no indication of a product quality issue with respect to labeling design or manufacturing of the device.
|
| |
|