|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
It was reported that during a gastric resection, the account performed full, complete squeezes of the handle, and tissue was cut without staples being deployed.The device was replaced with another device.
|
|
Additional information: d9, g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Functional testing found that the advanced staples in the sulu were reused and reset inside the cartridge.The instrument and the sulu were applied to appropriate test media.The jaw closed smoothly, the pin slide advanced fully, and the handle actuated smoothly into pre-clamped and clamped positions.The jaw opened, handle unlocked, and pin slide retracted when black release button was depressed.Acceptable staple formation was observed on test media.It was reported that the staples did not deploy.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.This issue may occur when the firing stroke is not completed during application.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.The instructions included with this device provide the following guidance: to fire the instrument, squeeze the handle a second time until it reaches a solid stop and it locks in the back position.If the ins trument handle is partially squeezed during firing then released, bleeding may occur as the instrument was not fully fired.The black release button may be used to open the instrument at any time during approximation.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|