The device was not returned for analysis.Production record and corrective and preventative actions (capa) reviews were performed and revealed no indication of a product quality issue.Additionally, a query of the complaint handling database for the reported lot revealed there is no indication of a lot specific issue.Based on the information received, the investigation determined that the reported balloon rupture appears to be related to operational context.There was no leak noted during preparation, which suggests a product quality issue did not contribute to the reported difficulties.In this case, it is likely that the balloon interacted with the moderately tortuous and moderately calcified anatomy resulting in the reported balloon rupture during inflation.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
|
|
It was reported that the procedure was to treat a 90% stenosed de novo lesion in the distal left circumflex (dlcx) artery with moderate calcification and moderate tortuosity.The 2.5x12mm nc trek neo balloon dilatation catheter (bdc) was advanced to the target lesion and the balloon was inflated; however, the balloon ruptured during the initial inflation at 6 atmospheres (atm).Therefore, the bdc was removed from the patient.A non-abbott bdc was used to continue the procedure.There was no adverse patient effect and no clinically significant delay reported in the procedure.No additional information was provided.
|