|
The investigation determined that a lower than expected tsh a single patient sample result was obtained when tested using the vitros immunodiagnostics products tsh reagent, lot 7590, on a vitros xt 7600 integrated system.This patient sample result was considered discordant when compared with repeat testing of the same sample.The assignable cause of the events was not determined with the information provided.Based on qc results obtained during the timeframe of the event, a reagent-related issue cannot be ruled out as a potential contributing factor.No precision testing was performed on the vitros xt 7600 integrated system at the time of the event; therefore, instrument performance could not be fully assessed, and an instrument-related issue cannot be ruled out as a potential contributor.In addition, pre-analytical sample processing cannot be excluded as a contributing factor.It was established that the customer was not following the sample collection device manufacturers centrifugation recommendations, and cellular debris due to inadequate sample preparation may have been present in the affected sample; however, this could not be confirmed.Furthermore, the time interval between the two testing events was not provided.Therefore, a patient sample mix-up or sample storage related issue cannot be ruled out as a potential contributor to the event.Continual tracking and trending of complaints has not identified any signals that would indicate a potential systematic issue with vitros tsh reagent lot 7590.
|
|
A customer contacted the ortho clinical diagnostics (ortho) technical solution center (tsc) to report a lower than expected tsh a single patient sample result was obtained when tested using the vitros immunodiagnostics products tsh reagent, lot 7590, on a vitros xt 7600 integrated system.This patient sample result was considered discordant when compared with repeat testing of the same sample.Vitros tsh patient result of <0.015 miu/l (hyperthyroid classification) vs.The expected result of 0.242 miu/l (hyperthyroid classification) the lower than expected vitros tsh result of <0.015 miu/l was not reported from the laboratory.There was no allegation of patient harm as a result of this event.This report corresponds to ortho clinical diagnostics inc (ortho) complaint number (b)(4) and reportability assessment 613627.
|