The manufacturer was contacted in reference to the voluntary field safety notice/recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.A remstar auto aflex device was returned to a manufacturer service center with no initial alleged complaint.During the evaluation, no foam particles were noted in the airpath; however, foam degradation was observed.The device¿s sound abatement foam was replaced to address the issue.The evaluation of the device data identified unrelated error codes.These error codes are consistent with normal device operation and expected use conditions and are not considered reportable under medical device reporting requirements.Additionally, no evidence was identified to suggest that these error codes contributed to or caused any event.The device was scrapped after the evaluation was completed.
|