• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON & CO (FRANKLIN LAKES) BD PRESET¿ ARTERIAL BLOOD COLLECTION SYRINGE; TUBES, VIALS, SYSTEMS, SERUM SEPARATORS, BLOOD COLLECTION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON & CO (FRANKLIN LAKES) BD PRESET¿ ARTERIAL BLOOD COLLECTION SYRINGE; TUBES, VIALS, SYSTEMS, SERUM SEPARATORS, BLOOD COLLECTION Back to Search Results
Catalog Number 364316
Medical Device Problem Code Device Contamination with Chemical or Other Material (2944)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/02/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation summary: bd received one sample and 3 photos for investigation.Evaluation of both shows evidence of foreign material inside the pouch.Additionally, ten retained samples were visually inspected, and no foreign material was found.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.This complaint has been confirmed for the indicated failure mode: foreign matter.No definitive root cause could be established for the reported defect.Based on the low defect rate for the batch in question, no actions are planned at this time.Complaints received for this device and reported condition will continue to be tracked and trended.Our business team regularly reviews the collected data for the identification of emerging trends.
 
Event or Problem Description
It was reported prior to using bd preset¿ arterial blood collection syringe, foreign matter was found in the sterile packaging of 1 device.No adverse impact was reported regarding the patient or user.
 
Event or Problem Description
It was reported prior to using bd preset¿ arterial blood collection syringe, foreign matter was found in the sterile packaging of 1 device.No adverse impact was reported regarding the patient or user.
 
Additional Manufacturer Narrative
The following fields were updated due to additional information: d9: device available for evaluation: yes, h3 device eval by manufacturer? yes, d9: returned to manufacturer on: 06-mar-2026.Investigation summary: bd received one sample and 3 photos for investigation.Evaluation of both shows evidence of foreign material inside the pouch.Additionally, ten retained samples were visually inspected, and no foreign material was found.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.This complaint has been confirmed for the indicated failure mode: foreign matter.No definitive root cause could be established for the reported defect.Complaints received for this device and reported condition will continue to be tracked and trended.Our business team regularly reviews the collected data for the identification of emerging trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD PRESET¿ ARTERIAL BLOOD COLLECTION SYRINGE
Common Device Name
TUBES, VIALS, SYSTEMS, SERUM SEPARATORS, BLOOD COLLECTION
Manufacturer (Section D)
BECTON DICKINSON & CO (FRANKLIN LAKES)
1 becton drive
franklin lakes NJ 07417
Manufacturer (Section G)
BECTON, DICKINSON AND COMPANY (BD)
belliver way
belliver industrial estate
plymouth
UK  
Manufacturer Contact
anand singh
1 becton drive
franklin lakes, NJ 07417
2018475186
MDR Report Key24503161
Report Number9617032-2026-00208
Device Sequence Number10486459
Product Code JKA
UDI-Device Identifier50382903643160
UDI-Public(01)50382903643160(11)250129(17)270131(10)5029026
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K202446
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/13/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue Number364316
Device Lot Number5029026
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/17/2026
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/02/2026
Supplement Date Received by Manufacturer03/13/2026
Initial Report FDA Received Date03/03/2026
Supplement Report FDA Received Date04/09/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/29/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-