| Model Number |
EPIQ 5 DIAMOND SELECT |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/03/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A customer reported their c10-3v intracavity transducer exhibited some damage to the lens tip area exposing a portion of the metal layer underneath.This probe is associated with the epiq 5 ultrasound system.There was no patient or user harm.A qualified service engineer evaluated the transducer to confirm the reported problem, and the transducer was replaced to address the customer's immediate concerns.Philips has started an investigation, and upon completion, a follow up report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
A thorough investigation was performed to determine the cause of the damages to the c10-3v transducer lens.Based on the evidence available in this investigation, the reported damage to the lens tip most closely aligns with use related factors such as repeated handling and frequent cleaning or sterilization cycles.There is no indication of non-conforming material at shipment, nor any evidence of a design anomaly that would warrant further action.
|
| |
|