Model Number EPIQ 5 DIAMOND SELECT
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/03/2026
Type of Reportable Event Malfunction
Event or Problem Description
A customer reported their c10-3v intracavity transducer exhibited some damage to the lens tip area exposing a portion of the metal layer underneath.This probe is associated with the epiq 5 ultrasound system.There was no patient or user harm.A qualified service engineer evaluated the transducer to confirm the reported problem, and the transducer was replaced to address the customer's immediate concerns.Philips has started an investigation, and upon completion, a follow up report will be submitted.
 
Additional Manufacturer Narrative
A thorough investigation was performed to determine the cause of the damages to the c10-3v transducer lens.Based on the evidence available in this investigation, the reported damage to the lens tip most closely aligns with use related factors such as repeated handling and frequent cleaning or sterilization cycles.There is no indication of non-conforming material at shipment, nor any evidence of a design anomaly that would warrant further action.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EPIQ 5
Common Device Name
SYSTEM, IMAGING, PULSED DOPPLER, ULTRASONIC
Manufacturer (Section D)
PHILIPS ULTRASOUND, INC
22100 bothell everett highway
bothell WA 98021
Manufacturer (Section G)
PHILIPS ULTRASOUND, INC
22100 bothell everett highway
bothell WA 98021
Manufacturer Contact
ondrea bermudez
22100 bothell everett highway
bothell, WA 98021
MDR Report Key24503556
Report Number3019216-2026-100072
Device Sequence Number10025481
Product Code IYN
UDI-Device Identifier00884838047686
UDI-Public(01)00884838047686
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K132304
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/21/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberEPIQ 5 DIAMOND SELECT
Device Catalogue Number795120
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/03/2026
Supplement Date Received by Manufacturer05/06/2026
Initial Report FDA Received Date03/03/2026
Supplement Report FDA Received Date05/21/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/24/2017
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
Yes
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-