There were multiple medical device types reported to be involved.The information for the additional device type is as follows: d.2.B medical device type: jka.D.2.A common device name: tubes, vials, systems, serum separators, blood collection.There were multiple 510k numbers reported to be involved.The information is as follows: g.4.Pma / 510(k)#: ;k213670;k231373.Investigation summary: bd received one sample along with three photos for investigation.Evaluation of the sample and photos revealed the presence of foreign matter under the label on the tube.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.This complaint has been confirmed for the indicated failure mode: foreign matter.No definitive root cause could be established for the reported defect.Based on the low defect rate for the batch in question, no actions are planned at this time.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.Our business team regularly reviews the collected data for the identification of emerging trends.
|