• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN EEA; STAPLER, SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN EEA; STAPLER, SURGICAL Back to Search Results
Model Number EEA28
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Failure to Anastomose (1028); Tissue Breakdown (2681)
Date of Event 02/26/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
It was reported that during use, after firing, the staple line was not holding steady.The device was replaced by a new circular stapler to resolve the issue.
 
Additional Manufacturer Narrative
H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted that the staple guide noted the instrument was fully applied.Further inspection noted that the green bar was visible in the indicator window and the safety feature was engaged.The staple guide was observed to be attached to the instrument with no damage to the staple guide.After actuating the handle, the pusher fingers appeared to be intact and inspection under the microscope displayed no damage to the knife blade.The anvil of the instrument was observed to be not tilted and attached to the instrument.Further inspection of the anvil cutting ring showed no traces of knife impression and the ring was received intact.Functional testing found that the wing nut and safety functioned properly upon rotating the twist knob.The green bar was visible within the indicator window when the twist knob was fully closed.The instrument was applied to the appropriate test media producing acceptable results.Pusher-fingers and knife advance when the handle was squeezed and the knife cut the media cleanly and completely.The device also produced all audible, tactile and visual indicators during the functional testing.It was reported that the of staple line did not hold.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.This issue may occur if the instrument handle compression is not completed or if the instrument is applied against an excessive amount of tissue during the clinical application.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: when firing the stapler, make sure to fully squeeze the instrument handle until the metal underside of the handle contacts the stapler body to the fullest extent.Partial or incomplete handle squeezes may result in unacceptable staple formation and/or incomplete knife cut.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EEA
Common Device Name
STAPLER, SURGICAL
Manufacturer (Section D)
COVIDIEN
sabanetas industrial park
ponce PR 00716 4401
Manufacturer (Section G)
COVIDIEN
sabanetas industrial park
ponce PR 00716 4401
Manufacturer Contact
mike bocchino
8200 coral sea st ne
mounds view, MN 55112-4391
8006464633
MDR Report Key24504252
Report Number2647580-2026-00582
Device Sequence Number18632045
Product Code GAG
UDI-Device Identifier10884523005530
UDI-Public(01)10884523005530(17)291031(10)P4L0775
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K221003
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/01/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberEEA28
Device Catalogue NumberEEA28
Device Lot NumberP4L0775
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/27/2026
Supplement Date Received by Manufacturer05/21/2026
Initial Report FDA Received Date03/04/2026
Supplement Report FDA Received Date06/01/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/18/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-