Model Number INTELLISPACE CARDIOVASCULAR
Medical Device Problem Code No Display/Image (1183)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/17/2026
Type of Reportable Event Malfunction
Event or Problem Description
Philips identified a software defect in iscv (intellispace cardiovascular) wherein during iscv (intellispace cardiovascular) installation, the data (sonography images, etc.) were deleted after just three days; studies were being stored in the repository only for 3 days.The device was in clinical use at the time of the event, and no adverse events or harm were reported in the complaint.Although no adverse events or patient harm were reported in the associated complaint (b)(4).This malfunction has been determined to be reportable.Under specific circumstances, it could potentially lead to delayed diagnosis or misdiagnosis.Given the defect¿s nature and its potential impact on clinical decision-making if it were to recur, philips has determined that this constitutes a reportable malfunction.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTELLISPACE CARDIOVASCULAR
Common Device Name
CARDIOVASCULAR IMAGE AND INFORMATION MANAGEMENT SYSTEM
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
veenpluis 6
best 5684 PC
NL  5684 PC
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
veenpluis 6
best 5684 PC
NL   5684 PC
Manufacturer Contact
edita reznik-shmueli
veenpluis 6
best 5684 -PC
NL   5684 PC
MDR Report Key24506065
Report Number3003768277-2026-100947
Device Sequence Number10440439
Product Code LLZ
UDI-Device Identifier00884838095250
UDI-Public(01)00884838095250(11)191207(10)4.2.1.0
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K153022
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 03/04/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberINTELLISPACE CARDIOVASCULAR
Device Catalogue Number830089
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/17/2026
Initial Report FDA Received Date03/04/2026
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device A
Patient Sequence Number1
Patient SexUnknown
-
-