| Model Number |
4473 |
| Medical Device Problem Code |
Low Readings (2460)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/03/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Other text: the masimo representative has confirmed the sensor will not return to allow an analysis to be performed.If the sensor is returned for evaluation or new information is obtained, a follow up report will be submitted.Health effect impact code: no adverse event; no patient impact.
|
| |
|
Event or Problem Description
|
|
The customer reported that "the spo2 value measured by mindray/masimo was "falsely" low compared with former philips/fast monitor with nellcor sensor and abg sao2: mindray/masimo - left hand: spo2 = 86; philips mp2/fast - foot: spo2 = 94; abg: sao2 = 92.4; 4 - 6 pts difference".There was no patient impact or consequence reported.
|
| |
|