THE ANSPACH EFFORT LLC EMAX 2 PLUS MOTOR; DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)
|
Back to Search Results |
|
| Catalog Number |
EMAX2PLUS |
| Medical Device Problem Codes |
Leak/Splash (1354); Vibration (1674)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H11 additional narrative: investigation summary: a visual and functional assessment was performed on the device: the following steps failed: visual assessment , loctite assessment, noise assessment, air pump assessment.During service/repair the following was observed (technician note): [pre-internal comment: ng, reproducible, connector looseness, hose puncture, housing pin looseness, abnormal noise.During the service evaluation the following failures were identified: functional : loose, visual : cord damaged, functional : noise - excessive, functional : vibration.Conclusion: failure reported is confirmed root cause: faulty parts.
|
| |
|
Event or Problem Description
|
|
It was reported that during service and evaluation, it was determined that the motor device had vibration.It was noted in the service order that the device had air leakage.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention, or prolonged hospitalization.The exact date of the event was unknown.However, it was reported that the event occurred in 2026.All available information has been disclosed.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11 additional narrative: a visual and functional assessment was performed on the device.The following steps failed: visual assessment, visual assessment, loctite assessment, noise assessment, air pump assessment.During the service evaluation the following failures were identified: functional : loose, visual : cord damaged, functional : noise - excessive, functional : vibration, functional : leak - air.Conclusion: failure reported is confirmed.
|
| |
|
Search Alerts/Recalls
|
|
|