• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. TWINFIX ULTRA HA 4.5 W/2 UB (WH & BLUE; FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. TWINFIX ULTRA HA 4.5 W/2 UB (WH & BLUE; FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE Back to Search Results
Catalog Number 72202597
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/17/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that, during a rotator cuff repair, a twinfix ultra ha 4.5 w/2 ub (wh & blue fractured during impact.The insertion site was cleared of all debris prior to insertion and the all the pieces were removed from the patient with suction.The procedure was resumed, without any delay, using an equivalent s+n back up used in the originally drilled bone hole.No further complications were reported and patient status is fine.
 
Additional Manufacturer Narrative
Internal complaint reference (b)(4).H11 h3, h6: this complaint was opened by smith+nephew to document a product problem associated with a smith+nephew device.The reported problem relates to known risks with the device itself or with its use that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are submitting this report in accordance with applicable regulations.If additional investigative findings or information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
 
Additional Manufacturer Narrative
H11: internal complaint reference: (b)(4).H3, h6: the reported device was not returned to the designated complaint unit for independent evaluation.However, an image evaluation was performed and found the anchor with a broken tip.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A review of the material specifications found that the storage requirements for the material are specified, and the material must comply with provided percentage composition specifications as measured by ash test.The root cause of the reported event could not be determined since findings cannot lead to a clear cause of the reported event.Factors that could have contributed to the reported event include an application of unintended inappropriate or excessive force to the device, attempted correction of a damaged device, or an impact event inconsistent with normal use.Based on this investigation, the need for corrective action is not indicated.
 
Additional Manufacturer Narrative
H11: internal complaint reference: (b)(4).H3, h6: the reported device was received for evaluation.A visual evaluation showed the device was returned in original packaging with the batch number of the complaint on the label.The sutures were not returned.The distal tip of the anchor has been fragmented.The rest of the anchor was not returned.The insertion device has no visible defect.Bio debris is present.An evaluation of the customer provided image showed the anchor with a broken tip.Based on the condition of the product material found during visual inspection, additional material testing is not required.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A review of the material specification found that the storage requirements for the material are specified, and the material must comply with provided percentage composition specifications as measured by ash test.A definitive root cause for the reported issue could not be determined.Factors that could have contributed to the failure include procedural variance, or deviations from established protocols, increasing risks and leading to unintended consequences (excessive force on the device, excessive torque on the device, attempted correction of a damaged device, off-axis insertion, improper preparation of the insertion site, or an inadvertent impact event inconsistent with normal use).Based on this investigation, the need for corrective action is not indicated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TWINFIX ULTRA HA 4.5 W/2 UB (WH & BLUE
Common Device Name
FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer (Section G)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer Contact
holly topping
1450 brooks rd.
memphis, TN 38116
5123913905
MDR Report Key24507518
Report Number1219602-2026-00696
Device Sequence Number21318047
Product Code MAI
UDI-Device Identifier03596010647559
UDI-Public3596010647559
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K151105
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 06/04/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number72202597
Device Lot Number2132541
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/11/2026
Supplement Date Received by Manufacturer04/10/2026
06/03/2026
Initial Report FDA Received Date03/04/2026
Supplement Report FDA Received Date04/13/2026
06/04/2026
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/19/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age65 YR
Patient SexFemale
Patient Weight57 KG
-
-