MAUDE Adverse Event Report: C.R. BARD INC. (COVINGTON) -1018233 BARDEX® LUBRI-SIL® ALL-SILICONE FOLEY CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD INC. (COVINGTON) -1018233 BARDEX® LUBRI-SIL® ALL-SILICONE FOLEY CATHETER Back to Search Results
Catalog Number 175818
Medical Device Problem Code Material Puncture/Hole (1504)
Health Effect - Clinical Code Discomfort (2330)
Date of Event 01/20/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that the three foley catheter from the same lot, all of which resulted in unexpected balloon deflation after insertion.Prior to insertion, nursing staff performed balloon integrity check and confirmed the retention balloon was intact and able to retain water.However, upon follow up assessment after balloon failure, a small hole was identified in the catheter, suggesting a potential defect not detectable during the pre-insertion balloon test.This defect required repeated catheter replacement and reinsertion, increasing patient discomfort and risk of complications.No medical intervention was reported.
 
Event or Problem Description
It was reported that indwelling foley catheters from lot ngkp1715 had presented a suspected catheter integrity defect that had affected product performance and patient safety.The issue had occurred with three (3) foley catheters from the same lot that had been inserted on (b)(6) 2026, (b)(6) 2026, and (b)(6) 2026, all of which had resulted in unexpected balloon deflation after insertion.Prior to insertion, nursing staff had performed a balloon integrity check and had confirmed that the retention balloon was intact and able to retain water.However, upon follow up assessment after the balloon failures, a small hole had been identified in the catheter, suggesting a potential defect that had not been detectable during the pre insertion balloon test.This defect had required repeated catheter replacement and reinsertion, increasing patient discomfort and the risk of complications.No medical intervention was reported.Reference: mw5183114, mw5183115 and mw5183116.
 
Additional Manufacturer Narrative
The reported event was inconclusive because this investigation did not result in any additional findings and no sample was available for evaluation the reported event is addressed within the labeling as it states: visually inspect the product for any imperfections or surface deterioration prior to use.If package is opened or any imperfection or surface deterioration is observed, do not use.Warning: on catheter, do not use ointments or lubricants having a petroleum base.They will damage the catheter and may cause balloon to burst.A review of the device history record did not show any problems or conditions that would have contributed to the reported issue.No additional actions can be taken at this time.Correction: d.Udi format updated with available product information per gudid.H11: sections a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDEX® LUBRI-SIL® ALL-SILICONE FOLEY CATHETER
Common Device Name
FOLEY CATHETER
Manufacturer (Section D)
C.R. BARD INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer (Section G)
C.R. BARD INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington, GA 30014
7707846100
MDR Report Key24508168
Report Number1018233-2026-01508
Device Sequence Number18637383
Product Code EZL
UDI-Device Identifier00801741034176
UDI-Public(01)00801741034176(11)250321(17)270630(10)NGKP1715
Combination Product (Y/N)N
Initial Reporter StatePR
Initial Reporter CountryUS
PMA/510(K) Number
K040504
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/31/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number175818
Device Lot NumberNGKP1715
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/26/2026
Supplement Date Received by Manufacturer03/31/2026
Initial Report FDA Received Date03/04/2026
Supplement Report FDA Received Date03/31/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/21/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexMale
-
-