| Model Number |
87480 |
| Medical Device Problem Codes |
Obstruction of Flow (2423); Material Deformation (2976)
|
| Health Effect - Clinical Codes |
Hypoventilation (1916); Low Oxygen Saturation (2477); Unintended Extubation (4564)
|
| Date of Event |
02/03/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that post-operatively, while patient was still under mechanical ventilation, airway emergency developed due to airway obstruction in endotracheal tube with inability to pass bronchoscopy through tube without foreign body obstruction.During the vid eolaryngoscopy, it was revealed a kink on the tube at the level of the oropharynx, at 17 cm marking with patient having difficult ventilation with tidal volume only 60-80 ml and desaturation occurred.Kinking was spontaneous, with no surrounding tissue swelling and adequate space around tube.Kinking remained despite manipulation and necessitated change of endo-tracheal tube.After exchange of tube, examination of the kinked tube revealed soft texture, easily compressed and kinked by manual manipulation.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: initial reporter and g3.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d9, g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.The evaluation found no potentially contributing factors, and the sample met all related specifications.It was reported that a kink on the tube at the level of the oropharynx , at 17 cm marking with patient having difficult ventilation with tidal volume only 60-80 ml and desaturation occurred.Kinking was spontaneous, with no surrounding tissue swelling and adequate space around tube.Kinking remained despite manipulation and necessitated change of endo-tracheal tube.The reported issue could not be confirmed.The most likely cause could not be identified because no related problem was detected with the device.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|