Model Number DVPB3D4
Medical Device Problem Codes Signal Artifact/Noise (1036); Inappropriate/Inadequate Shock/Stimulation (1574)
Health Effect - Clinical Code Shock from Patient Lead(s) (3162)
Date of Event 09/22/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that during the replacement procedure, the implantable cardioverter defibrillator (icd) had problems because part of the right ventricular (rv) lead was stuck inside.It was also noted that the pin of the lead had came off.A different device was successfully implanted and the lead was capped and replaced.It was further reported by the patient that they had initially heard alert tones and didn¿t realize they were coming from their implanted device.The patient reported that when they were out of state, they received a shock and then after that, if they moved their neck a certain way, they got more shocks, receiving four more.Once back home, the patient saw their heart doctor and was told there was noise on the icd and it needed to be replaced.When replacing the device, the surgery took longer as the doctor said a lead was corroded and was broke in half.Later that evening, a lead came loose from the new icd post-surgery, and the patient had surgery the following morning to reattach the lead.The patient said the intravenous (iv) placed had an occlusion and they were never ¿out¿ for the surgery.The patient stated they got a blood clot and were on medication for it.The new icd and lead remain in use.No further patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
It was further reported by the patient that the original lead had corroded and broke in have and that is why the patient was shocked.Additionally, the patient indicated during the surgery the patient had an iv infiltration and left arm swelled enormously and ended up with a deep vein thrombosis (dvt) from the carotid artery to the elbow.
 
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COBALT¿ VR MRI SURESCAN¿
Common Device Name
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT)
Manufacturer (Section D)
MEDTRONIC EUROPE SARL
route du molliau 31
case postale
tolochenaz vaud 1131
SZ  1131
Manufacturer (Section G)
MEDTRONIC EUROPE SARL
route du molliau 31
case postale
tolochenaz vaud 1131
SZ   1131
Manufacturer Contact
gerard torenvliet
8200 coral sea st ne
mounds view, MN 55112--439
7635269277
MDR Report Key24511473
Report Number9614453-2026-00874
Device Sequence Number18627771
Product Code LWS
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P980016
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/07/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date12/28/2025
Device Model NumberDVPB3D4
Device Catalogue NumberDVPB3D4
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/10/2026
Supplement Date Received by Manufacturer06/17/2026
Initial Report FDA Received Date03/04/2026
Supplement Report FDA Received Date07/07/2026
Date Device Manufactured07/09/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
6946M62 LEAD
Outcome Attributed to Adverse Event Required Intervention; Life Threatening; Hospitalization;
Patient Age64 YR
Patient SexFemale
-
-