• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CAREFUSION, INC BD SURGIPHOR¿ STERILE WOUND IRRIGATION SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CAREFUSION, INC BD SURGIPHOR¿ STERILE WOUND IRRIGATION SYSTEM Back to Search Results
Catalog Number 910110
Medical Device Problem Code Crack (1135)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/03/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
As reported, it is alleged that the two surgiphor bottles (device 1 and 2) cracked during use.It was also reported that the representative present with the surgeon indicated that the bottles may have been opened incorrectly, which required him to apply excessive squeezing.There was no patient or user injury reported.This mdr represents surgiphor bottle (device #1).Additional mdr submitted to represent surgiphor bottle (device #2) section c.1a name and strength: 0.5% povidone iodine (pvp-i) in phosphate-buffered saline, potassium iodide, and vitamin e tpgs section d.2b procode: fqh (lavage, jet) and fro (dressing, wound, drug) note: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
As reported, it is alleged that the two surgiphor bottles (device 1 and 2) cracked during use.It was also reported that the representative present with the surgeon indicated that the bottles may have been opened incorrectly, which required him to apply excessive squeezing.There was no patient or user injury reported.
 
Additional Manufacturer Narrative
As reported, it is alleged that the two surgiphor bottles (device 1 and 2) cracked during use.It was also reported that the representative present with the surgeon indicated that the bottles may have been opened incorrectly, which required him to apply excessive squeezing.There was no patient or user injury reported.Section c.1a name and strength: 0.5% povidone iodine (pvp-i) in phosphate-buffered saline, potassium iodide, and vitamin e tpgs.Section d.2b procode: fqh (lavage, jet) and fro (dressing, wound, drug).Addendum: this supplemental mdr is submitted to document the results of investigation performed to analyze the root cause.Based on the information available and without having the photos or sample to evaluate, no conclusions can be made.A device history record review found no non-conformances associated with this issue during production of the reported batch.Additionally, retain samples were inspected and found no visible abnormalities or signs of the reported cracking on the bottle.A capa was opened to investigate events of this nature.Complaints are monitored and reviewed for emerging trends.Updated fields: b4, g3, g6, h2, h6, h10.This supplemental mdr represents surgiphor bottle (device #1).Additional supplemental mdr submitted to represent surgiphor bottle (device #2).Note: section a through f: the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
As reported, it is alleged that the two surgiphor bottles (device 1 and 2) cracked during use.It was also reported that the representative present with the surgeon indicated that the bottles may have been opened incorrectly, which required him to apply excessive squeezing.There was no patient or user injury reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD SURGIPHOR¿ STERILE WOUND IRRIGATION SYSTEM
Common Device Name
SURGIPHOR
Manufacturer (Section D)
CAREFUSION, INC
75 n. fairview drive
vernon hills IL 60061
Manufacturer (Section G)
CAREFUSION, INC
75 n. fairview drive
vernon hills IL 60061
Manufacturer Contact
andrew topoulos
75 north fairway drive
vernon hills, IL 60061
8003239088
MDR Report Key24512098
Report Number1423507-2026-00036
Device Sequence Number9960011
Product Code FQH
UDI-Device Identifier00382909101102
UDI-Public(01)00382909101102(17)270228(10)5050076
Combination Product (Y/N)Y
Initial Reporter StateUT
Initial Reporter CountryUS
PMA/510(K) Number
K221504
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/24/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number910110
Device Lot Number5050076
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/04/2026
Supplement Date Received by Manufacturer03/24/2026
Initial Report FDA Received Date03/04/2026
Supplement Report FDA Received Date04/14/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/10/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-