ABBOTT MEDICAL SWARTZ¿ BRAIDED TRANSSEPTAL GUIDING INTRODUCER, SL1¿, 63 CM LENGTH, 8.5F; Introducer, catheter
|
Back to Search Results |
|
| Model Number |
407453 |
| Medical Device Problem Code |
Gas/Air Leak (2946)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/04/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
One 8.5f swartz braided introducer sheath was received for evaluation.Functional testing determined a leak was noted in the hemostasis valve.The cap was removed from the hemostasis hub and the hemostasis seals were microscopically inspected.Tearing, resulting in a hole, was noted in the proximal and distal seals.The batch record was reviewed to ensure that each manufacturing and inspection operation was performed.The ifu states: do not remove dilator or catheter rapidly.Damage to the valve may occur, potentially compromising hemostasis.
|
| |
|
Event or Problem Description
|
|
During the atrial fibrillation procedure, air was aspirated from the hemostasis valve.The introducer was replaced, but the same issue occurred.The introducer was replaced again, and the procedure was completed with no adverse patient consequences.
|
| |
|