BIOSENSE WEBSTER INC OPTRELL MAPPING CATHETER WITH TRUEREF TECHNOLOGY; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING
|
Back to Search Results |
|
| Catalog Number |
D142901 |
| Medical Device Problem Codes |
Adverse Event Without Identified Device or Use Problem (2993); Patient Device Interaction Problem (4001)
|
| Health Effect - Clinical Codes |
Cardiac Tamponade (2226); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/26/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Since no device has been received for analysis, no product investigation can be performed, and the customer complaint cannot be confirmed.Investigation is in progress, once completed a supplemental will be submitted.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster, inc., or its employees that the report constitutes an admission that the product, biosense webster, inc., or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Manufacturer's reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
During a clinical trial it was reported that a patient underwent a ventricular tachycardia procedure with a qdot micro catheter and an optrell mapping catheter with trueref technology and suffered a pericardial effusion which required prolonged hospitalization.Patient received an index ventricular tachycardia cardiac ablation on (b)(6) 2026.On (b)(6) 2026 start date and site awareness date of 06-feb-2026, the patient experienced pericardial effusion without hemodynamic relevance after the ablation procedure with a qdot micro catheter categorized as moderate and serious defined by in patient/prolonged hospitalization with an admission date of (b)(6) 2026 and discharged date of (b)(6) 2026.Relationship to study devices (qdot micro and optrell mapping catheter with trueref technology) was probable and not related to all other devices.The relationship with the index study procedure is causal and expected/anticipated.The outcome is resolved with an end date of (b)(6) 2026 with no intervention noted.The date the event is first reported to be an serious adverse event (sae) is 06-feb-2026.
|
| |
|
Additional Manufacturer Narrative
|
|
Initially the adverse event was reported as mdr reportable under both the qdot micro catheter (manufacturer report number: 2029046-2026-00703) and an optrell mapping catheter with trueref technology (manufacturer report number: 2029046-2026-00713).However, during an internal review on (b)(6) 2026, the event was reassessed as only mdr reportable under the qdot micro catheter as this device delivered the energy and was most likely the device associated with the adverse event.Therefore, the optrell mapping catheter with trueref technology was reassessed as not mdr reportable.Hence, corrected the h6.Health effect - clinical code, h6.Health effect - impact code and h6.Medical device problem code fields.Manufacturer's reference number: (b)(4).
|
| |
|
Search Alerts/Recalls
|
|
|