• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AXONICS MODULATION TECHNOLOGIES, INC. AXONICS; STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCONTINENCE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AXONICS MODULATION TECHNOLOGIES, INC. AXONICS; STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCONTINENCE Back to Search Results
Model Number 1101
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/01/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Block d2b: product code - additional product code qon.Block g4: premarket / 510(k) # - additional premarket/510(k) p190006.Block h11: udi for concomitant product, tined lead: (b)(4).
 
Event or Problem Description
It was reported that the patient no longer has their axonics system.The patient stated it did not work for them.There were no other issues or complications reported.
 
Additional Manufacturer Narrative
Block d2b: product code - additional product code qon block c2/d10: model# 1201 sn# (b)(6).Model: tined lead udi# (b)(4).Block g4: premarket / 510(k) # - additional premarket/510(k) p190006 block h11: as the complaint component was not returned for analysis, and the product record review revealed no additional information related to the complaint, the reported allegation could not be confirmed.The event cannot be reproduced or substantiated.It was confirmed the device met manufacturing specification prior to distribution and there were no manufacturing deviations which could have contributed to the reported event.Review of the instructions for use (ifu) confirmed there was relevant content and sufficient guidance with respect to the circumstances described within this complaint.No updates are required to the ifu as a result of this event.A risk review was completed and confirmed that the events of inadequate/inappropriate treatment or diagnostic exposure and device explantation were defined in the product risk documentation.This event type has been accounted for during analysis to support acceptable risk benefit for the product.Based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of known inherent risk of device.
 
Event or Problem Description
It was reported that the patient no longer has their axonics system.The patient stated it did not work for them.There were no other issues or complications reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AXONICS
Common Device Name
STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCONTINENCE
Manufacturer (Section D)
AXONICS MODULATION TECHNOLOGIES, INC.
15515 sand canyon ave
irvine CA 92618
Manufacturer (Section G)
AXONICS INC
26 technology dr
irvine CA 92618
Manufacturer Contact
farshad fahimi
4100 hamline avenue north
building c
saint paul, MN 55112
MDR Report Key24514535
Report Number2124215-2026-12150
Device Sequence Number18578506
Product Code EZW
UDI-Device Identifier10810005340066
UDI-Public(01)10810005340066(11)210422(17)240422
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P180046
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2026
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/21/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date04/22/2024
Device Model Number1101
Device Catalogue Number1101
Device Lot NumberAX1G171280
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/05/2026
Supplement Date Received by Manufacturer03/23/2026
Initial Report FDA Received Date03/04/2026
Supplement Report FDA Received Date04/21/2026
Date Device Manufactured04/22/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
TINED LEAD (1201) SN: (B)(6).
Outcome Attributed to Adverse Event Other; Hospitalization;
Patient Age75 YR
Patient SexFemale
-
-