| Model Number |
REF 316-08 |
| Medical Device Problem Code |
Loss of or Failure to Bond (1068)
|
| Health Effect - Clinical Codes |
Aspiration/Inhalation (1725); Low Oxygen Saturation (2477)
|
| Date of Event |
02/11/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported on 2026-02-11 that subglottic port became detached from tracheostomy tube.The patient is very dependant on sgp suction and ventilated overnight, required an emergency tracheostomy change.The patient took a long time to optimise post cuff deflation due to burden of secretions, causing discomfort and distress and drop in saturations.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation of the affected sample was not possible because the product was not returned to tracoe.No photos of defective device was available.Based on the incident description it can be supposed that the defect described relates to the known product error of the subglottic suction line detachment.Further investigations are conducted within capa 000106/330.The cause is probably due to the bonding process of the subglottic suction line.It is possible that preparation steps before bonding were not carried out sufficiently or correctly, or that the bonding was not carried out correctly.Further root cause can be attributed to a design limitation of the component.Other causes could also be seen within the use period, e.G.The use of different cleaning or disinfection agents, humification with a humification mask, as well as strong pulls on the suction line.
|
| |
|