| Model Number |
1888TC/46 |
| Medical Device Problem Code |
No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
02/27/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
This report is to advise of an event observed during analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
The lead was returned with no reported event.As received only the connector portion of the lead was returned in one piece.Visual examination found full breach outer insulation degradation adjacent to the connector boot exposing the outer coil.All the other damage noted is consistent with procedural damage.
|
| |
|