On 2025-dec-26 mpxr 1379023 (rep, hcp, for): information was received from healthcare provider (hcp) via a manufacturer representative regarding an event which occurred prior to use.It was reported that the screw is not fixing. there was no patient involvement reported.
|
|
H3: product analysis: product: 7578302, lot: em24f005 visual examination identified that one of the gripper retaining tabs was bent, thus preventing the instrument from being able to fix the screw.The damage is consistent with bend-stress overload.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|