• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ICU MEDICAL HEALTHCARE MANUFACTURING S.A. DE C.V. LEVEL 1 LOW FLOW HOTLINE DISPOSABLE SET; WARMER, THERMAL, INFUSION FLUID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ICU MEDICAL HEALTHCARE MANUFACTURING S.A. DE C.V. LEVEL 1 LOW FLOW HOTLINE DISPOSABLE SET; WARMER, THERMAL, INFUSION FLUID Back to Search Results
Catalog Number L-70
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/01/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
One used hotline warming set-disp was returned for evaluation.Upon visual inspection, a hairline cracks were observed on the upstream luer of the infusate fluid path.Upon functional testing, an attempt to prime the infusate line was done using an icu medical provided syringe.A leak was observed originating from the crack on the infusate line luer.The complaint of leakage can be confirmed.The probable cause is due to unintended external forces on the luer.The lot history was reviewed; no nonconformities were identified that may have contributed to the reported complaint.Date of event: since the exact event date was unknown in december, it was updated as first day of december.
 
Event or Problem Description
It was found during investigation of the returned device that a hairline crack and leak was found on the on the infusate line luer.There was patient involvement, but no patient harm.This report captures the second of two occurrences.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 LOW FLOW HOTLINE DISPOSABLE SET
Common Device Name
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
ICU MEDICAL HEALTHCARE MANUFACTURING S.A. DE C.V.
ave calidad no. 4, parque
tijuana
MX 
Manufacturer Contact
reed covert
6000 nathan lane north
minneapolis, MN 55442
2247062300
MDR Report Key24522421
Report Number3012307300-2026-02110
Device Sequence Number18271372
Product Code LGZ
UDI-Device Identifier30695085407007
UDI-Public(01)30695085407007(17)270127(10)4364734
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K911383
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 03/05/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberL-70
Device Lot Number4364734
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/31/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/12/2026
Initial Report FDA Received Date03/05/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/14/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-