Catalog Number 55-10554
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/17/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a plate fracture was identified, and revision surgery is scheduled.
 
Additional Manufacturer Narrative
Device is not available for evaluation.If additional information is received, it will be reported on a supplemental report.
 
Event or Problem Description
No new information.
 
Additional Manufacturer Narrative
Additional information: h6.Correction: d2a, d9.The reported plate breakages were confirmed based on the provided x rays.Additional information indicated that the plates and screws used were components of the universal 2.0 system and intended to be used together.Medical review concluded that the plates were positioned and fixated properly.Based on the available information, no evidence of a device related, manufacturing, or design issue was identified.Based on the investigation, there is no indication for an incorrectly working product or any design, material, or manufacturing related issue.Therefore, no corrective and/or preventive actions are deemed necessary at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PLT/6HL CVD SAG SPT,6MMBAR LOK
Common Device Name
PLATE, BONE
Manufacturer (Section D)
STRYKER LEIBINGER FREIBURG
boetzingerstr. 41
freiburg D-791 11
GM  D-79111
Manufacturer (Section G)
Stryker Leibinger Freiburg
boetzingerstr. 41
freiburg D-791 11
GM   D-79111
Manufacturer Contact
elena gehring
boetzingerstr. 41
freiburg D-791-11
GM   D-79111
+49()7614888208
MDR Report Key24523649
Report Number0008010177-2026-00028
Device Sequence Number9704167
Product Code JEY
UDI-Device Identifier04546540366245
UDI-Public04546540366245
Combination Product (Y/N)N
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial,Followup
Report Date (Section B) 09/14/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Catalogue Number55-10554
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/18/2026
Supplement Date Received by Manufacturer02/18/2026
Initial Report FDA Received Date03/05/2026
Supplement Report FDA Received Date09/14/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-