| Model Number |
M635WU60200 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Low Blood Pressure/ Hypotension (1914); Cardiac Tamponade (2226); Pericardial Effusion (3271)
|
| Date of Event |
03/02/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported a pericardial effusion occurred.A left atrial appendage closure (laac) procedure was being performed.Prior to the procedure, a trivial pericardial effusion was noted at baseline on imaging.A watchman access system (was) and a 20mm watchman flx pro laa closure device & delivery system (wds) were selected for use.The transseptal puncture (tsp) was performed and was noted to be difficult due to a thick septum.The 20mm wds was deployed in the laa.Imaging was difficult to view in the mitral view.While repositioning for a higher angle, it was noted that there was fluid around the left ventricle (lv).The patient blood pressure was checked and noted to be 65 systolic.A pericardial tap kit was prepared, and a transthoracic echocardiogram (tte) was taken.A pericardial effusion was noted on both the lv and right ventricle (rv).Approximately 1l of blood was drained to treat the effusion.The patient stabilized after 60 ml protamine was administered and 1000ml of blood was drained.The blood was transfused back into the patient.Tte showed improvement of the effusion and the patient blood pressure stabilized at 120 systolic.The patient was monitored for 15 minutes.Transesophageal echocardiography (tee) confirmed the closure device was in a stable position and meeting position, anchor, size and seal (pass) criteria.The closure device was released.The patient was sent to the intensive care unit (icu) for recovery and was scheduled to receive an additional two units of blood later.There were no further complications and the patient was discharged.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for analysis as the device remains implanted; therefore, a technical analysis could not be performed.It was confirmed this device met manufacturing specification prior to distribution and there were no manufacturing deviations which could have contributed to the reported event.Based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of known inherent risk of device.
|
| |
|
Event or Problem Description
|
|
It was reported a pericardial effusion occurred.A left atrial appendage closure (laac) procedure was being performed.Prior to the procedure, a trivial pericardial effusion was noted at baseline on imaging.A watchman access system (was) and a 20mm watchman flx pro laa closure device & delivery system (wds) were selected for use.The transseptal puncture (tsp) was performed and was noted to be difficult due to a thick septum.The 20mm wds was deployed in the laa.Imaging was difficult to view in the mitral view.While repositioning for a higher angle, it was noted that there was fluid around the left ventricle (lv).The patient blood pressure was checked and noted to be 65 systolic.A pericardial tap kit was prepared, and a transthoracic echocardiogram (tte) was taken.A pericardial effusion was noted on both the lv and right ventricle (rv).Approximately 1l of blood was drained to treat the effusion.The patient stabilized after 60 ml protamine was administered and 1000ml of blood was drained.The blood was transfused back into the patient.Tte showed improvement of the effusion and the patient blood pressure stabilized at 120 systolic.The patient was monitored for 15 minutes.Transesophageal echocardiography (tee) confirmed the closure device was in a stable position and meeting position, anchor, size and seal (pass) criteria.The closure device was released.The patient was sent to the intensive care unit (icu) for recovery and was scheduled to receive an additional two units of blood later.There were no further complications and the patient was discharged.
|
| |
|