| Model Number |
MODEL 200 |
| Medical Device Problem Code |
Use of Device Problem (1670)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/11/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported that during a demo, they placed a torso manikin on the autopulse nxt platform (sn (b)(6)).The platform and nxt band (lot # unknown) kept providing compressions; however, the band twisted on its own.No patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
For g4: pma/510(k): premarket submission number not available/not released.Zoll has not received the autopulse nxt platform in complaint for investigation.A follow-up report will be filed if and when the product is returned and investigation has been completed.
|
| |
|
Event or Problem Description
|
|
The customer reported that during a demonstration, a torso manikin was positioned in the center of the autopulse nxt platform (sn (b)(6), where the patient would be placed.The nxt band (lot #211836) was installed across the manikin's chest in the appropriate anatomical position.When the nxt platform started compressions, the band twisted on its own.The customer reported that the band initially took up on only one side.No alerts were displayed on the nxt platform during the event.The band had not been used in any prior demonstrations, and no visible damage or signs of deterioration were observed.When the same torso manikin was used with different nxt band(s), the nxt platform functioned as expected.No patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
Sections b5, d9, h3, h4, and h6 codes were updated.The reported complaint that the nxt band twisted on its own after the initiation of compressions with the autopulse nxt platform (sn (b)(6) was not confirmed during functional testing at zoll.The autopulse nxt platform functioned appropriately and performed as intended.Based on the archive log review, no fault or alert occurred during the event.The autopulse nxt platform successfully passed preliminary functional testing with no faults or errors.The nxt platform was tested using a zoll test band with the medium resuscitation test fixture (mrtf) until the battery was fully discharged.During testing, no band twisting was observed.The nxt platform successfully passed all testing with no issues.Following service, the autopulse nxt platform passed all testing criteria.
|
| |
|