|
Block e1: this event was reported by the distributor.The healthcare facility where the event occurred is: (b)6) hospital.Phone number: (b)(6), fax number: (b)(6).Block h6: imdrf device code a0402 captures the reportable event of balloon burst block h11: investigation results the returned cre rx dilatation balloon was analyzed, and a visual inspection found no damages to the device.Functional inspection was performed, and the balloon was able to be inflated without any problem.However, it was not able to hold the pressure due to it had a pinhole in the balloon which produces a failure to inflate, located approximately at 20 mm from the tip of the device.Microscopic inspection confirmed the balloon pinhole.No other problems with the device were noted.A review of the manufacturing documentation for this device revealed that no anomalies or deviations related to the event occurred during manufacturing.With all the available information, boston scientific concludes the reported event of balloon burst was not confirmed.The pinhole problem found could have been interpreted by the customer as the reported event of a balloon burst.The pinhole found in the balloon is likely to have occurred due to procedural factors such as excess of pressure, interaction with other devices, or anatomical conditions.Also, it is possible that interaction with a sharp surface during or previous to the procedure, could have caused the problem found on the distal section of the balloon.Therefore, the most probable root cause is an adverse event related to procedure.
|