| Model Number |
MI1250 SYNCHRONY 2 |
| Medical Device Problem Code |
Migration (4003)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/03/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Two electrode channels migrated out of cochlea after implantation.Revision surgery was performed on (b)(6) 2026 and the user was implanted with a new device.
|
| |
|
Additional Manufacturer Narrative
|
|
The device has been explanted and should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
|
| |
|
Additional Manufacturer Narrative
|
|
Conclusion: device investigation did not reveal any device defect or damage, which has been present whilst implanted.Mechanical damages found are attributable to the removal surgery.According to the information received from the field the device was explanted due to a post-operative migration of the active electrode out of cochlea.This is a final report.
|
| |
|
Event or Problem Description
|
|
Two electrode channels migrated out of cochlea after implantation.Revision surgery was performed on (b)(6) 2026 and the user was implanted with a new device.At explantation 6 electrode channels were found extra-cochlear.
|
| |
|