• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOTRONIK AG ORSIRO MISSION US, 3.0/35; CORONARY DRUG-ELUTING STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOTRONIK AG ORSIRO MISSION US, 3.0/35; CORONARY DRUG-ELUTING STENT Back to Search Results
Model Number 453970
Medical Device Problem Code Failure to Advance (2524)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/12/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Combination product: yes.Neither the affected device nor the angiographic material was returned.Therefore, no technical investigation on the subject could be performed.The production documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for the reported event.Review of the production documentation confirmed that the device was manufactured according to specifications and passed all in-process and final inspections.Based on the conducted investigations, no material or manufacturing related root cause could be determined.
 
Event or Problem Description
The orsiro mission drug-eluting stent system was selected for treatment of a moderately calcified lesion (95 % stenosis degree) in a moderately tortuous part of the mid lcx.The lesion was pre-dilated, and the affected device was introduced into the patient's body but could not cross the lesion.There was no harm to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ORSIRO MISSION US, 3.0/35
Common Device Name
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
BIOTRONIK AG
ackerstrasse 6
bulach, zurich 8180
SZ  8180
Manufacturer (Section G)
BIOTRONIK AG
ackerstrasse 6
bulach, zurich 8180
SZ   8180
Manufacturer Contact
jana burger
ackerstrasse 6
bulach, zurich 8180
SZ   8180
MDR Report Key24529524
Report Number8043892-2026-02131
Device Sequence Number8933600
Product Code NIQ
UDI-Device Identifier07640130456266
UDI-Public(01)07640130456266(17)280826(10)07254471
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P170030
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 03/06/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number453970
Device Lot Number07254471
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/13/2026
Initial Report FDA Received Date03/06/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/28/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-