| Catalog Number |
09793194001 |
| Medical Device Problem Codes |
High Test Results (2457); Non Reproducible Results (4029)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/17/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The analyzer's serial number is (b)(6) a precision test was performed with acceptable results.The investigation is ongoing.
|
| |
|
Event or Problem Description
|
|
There was an allegation of questionable potassium electrode results for 1 patient sample on a cobas c 303 analytical unit.The initial result was approximately 7.0 mmol/l.The repeated result was approximately 3.5 mmol/l.The repeated result was believed to be correct.The sample was repeated because the initial result was marked as critical.
|
| |
|
Additional Manufacturer Narrative
|
|
Sections d1-d4, g1, and g4 were updated.The calibration and qc were acceptable.The alarm trace showed "abnormal aspiration (sample pipetter)" errors.The field service representative did not identify a root cause for the issue.The electrodes were replaced, and the field service representative verified that the instrument was performing within specification.The investigation could not determine a specific cause of the event.Although no cause was identified, the issue appears to be resolved.The investigation did not identify a product problem.
|
| |
|
Search Alerts/Recalls
|