MAUDE Adverse Event Report: SYNTHES GMBH VA-LCP DHP 2.7/3.5 DORSO-LAT R SHORT 3HO; PLATE, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH VA-LCP DHP 2.7/3.5 DORSO-LAT R SHORT 3HO; PLATE, FIXATION, BONE Back to Search Results
Catalog Number 02.117.203
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Nonunion/Delayed-union Bone Fracture (2369); Malunion of Bone (4529)
Date of Event 02/23/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported on (b)(6) 2026, that the patient was revised and hardware was removed due to non-union and mal-union.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.D4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: added: d10 if information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11.Additional narrative: h6.Component codes: most relevant component code is g07002 (appropriate term/code not available) to capture no findings available due to no product returned.H6: the product was not returned to depuy synthes; however, photos were provided for review.The photo and x-ray investigation revealed nothing indicative of a device nonconformance.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.The overall complaint was not confirmed as the observed condition of the va-lcp dhp 2.7/3.5 dorso-lat r short 3ho would have not contributed to the reported adverse event.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: the device lot number is unknown; therefore, a device history review could not be performed.If more information become available, the record will be re-assessed.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VA-LCP DHP 2.7/3.5 DORSO-LAT R SHORT 3HO
Common Device Name
PLATE, FIXATION, BONE
Manufacturer (Section D)
SYNTHES GMBH
luzernstrasse 19-21
zuchwil 4528
SZ  4528
Manufacturer (Section G)
MONUMENT
1101 synthes avenue
monument CO 80132
Manufacturer Contact
elba bello
1302 wright lane east
west chester, PA 19380
9083863534
MDR Report Key24530241
Report Number8030965-2026-02260
Device Sequence Number10337676
Product Code HRS
UDI-Device Identifier10886982036100
UDI-Public(01)10886982036100
Combination Product (Y/N)N
Initial Reporter StateUT
Initial Reporter CountryUS
PMA/510(K) Number
K120717
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number02.117.203
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/23/2026
Supplement Date Received by Manufacturer03/11/2026
03/30/2026
Initial Report FDA Received Date03/06/2026
Supplement Report FDA Received Date03/17/2026
03/31/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNK - SCREWS: 2.7 MM VA LOCKING; UNK - SCREWS: 2.7 MM VA LOCKING; UNK - SCREWS: 2.7 MM VA LOCKING; UNK - SCREWS: 2.7 MM VA LOCKING; UNK - SCREWS: 2.7 MM VA LOCKING; UNK - SCREWS: 2.7 MM VA LOCKING; UNK - SCREWS: 2.7 MM VA LOCKING; UNK - SCREWS: 2.7 MM VA LOCKING; UNK - SCREWS: 3.5 MM CORTEX; UNK - SCREWS: 3.5 MM CORTEX; UNK - SCREWS: 3.5 MM CORTEX; UNK - SCREWS: 3.5 MM LOCKING; UNK - SCREWS: METAPHYSEAL; VA-LCP DHP 2.7/3.5 MED R MED 2HO L82 SST
Outcome Attributed to Adverse Event Required Intervention;
Patient Age61 YR
Patient SexFemale
-
-