| Model Number |
BF-1T180 |
| Medical Device Problem Code |
Thermal Decomposition of Device (1071)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/12/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported during reprocessing, the plastic distal end cover of the bronchovideoscope was burned.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation is ongoing.A supplemental report will be submitted when the investigation is completed or if additional information becomes available.
|
| |
|
Event or Problem Description
|
|
No additional information received from customer.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental report is being submitted to provide the results of the legal manufacturer's investigation.The device was returned to olympus for inspection and the reported event was confirmed.A root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: a high-energy treatment tool (high frequency, etc.) was used without ensuring a sufficient distance from the tip.Additionally, the distal end was corroded most likely due to stress.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|