| Catalog Number |
8065751761 |
| Medical Device Problem Code |
Suction Problem (2170)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/12/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A healthcare professional reported that during a cataract surgery, an ophthalmic handpiece failed to aspirate in the phacoemulsification mode.The surgery was completed on the same day using a new handpiece, and there was no patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
The product under investigation is not a serviceable device.Therefore, a service record review was not performed.However, a handpiece (hp) was returned for testing on this investigation.A non-conformance based review of the serial number was performed and did not reveal any potential contributing factors to the reported complaint.A manufacturing device history record (dhr) review was performed prior to product release to ensure that the product was manufactured in compliance with the device master record.Based on the assessment, the product met release criteria.A review for ¿complaints reported against this serial number¿ was performed.There were no similar complaints reported for the ¿product serial¿ under investigation, with the same event code.The hp was received for testing on this investigation.A visual assessment of the returned sample found discolored red dot.A flow rate test was performed on the irrigation and aspiration lines of the handpiece which found the handpiece to meet specifications.The handpiece was connected to a calibrated resistance test box, where resistance and dissipation were found to be within specification.The handpiece was connected to dynamic tuning fixture (dtf) for tune data which found the handpiece to not meet alcon specifications.Disassembling the handpiece found broken wires on the side of the handpiece connector.The returned handpiece was found to functionally exhibit no problems.Therefore, the root cause of the reported event is inconclusive.Manufacturer will continue to monitor data for evidence of adverse trending and take further action, as appropriate.Similar incident/event data was collected for the associated reported events.Quality assurance will continue to perform periodic monitoring for evidence of adverse trending and take further action as appropriate.The manufacturer internal reference number is: (b)(4).
|
| |
|
Search Alerts/Recalls
|