| Model Number |
9-ASD-026 |
| Medical Device Problem Code |
Difficult to Fold, Unfold or Collapse (1254)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/09/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2026, a 26mm amplatzer septal occluder was chosen for implant using a 10f amplatzer trevisio delivery system.While deployment noted that the device was not taking the proper shape.They tried multiple times but the results remain the same.So, they removed the device back.
|
| |
|
Additional Manufacturer Narrative
|
|
The device is expected to be returned for investigation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.
|
| |
|
Additional Manufacturer Narrative
|
|
An event of device deformity could not be confirmed.The device was returned to abbott for investigation and the device met functional specifications when analyzed under non-physiological conditions.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.There were no complaints associated with any other devices from the lot.The cause of the reported incident could not be conclusively determined.There is no indication of a product quality issue with respect to labeling design or manufacturing of the device.
|
| |
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2026, a 26mm amplatzer septal occluder was chosen for implant using a 10f amplatzer trevisio delivery system.While deployment noted that the device was not taking the proper shape and had a cobra deformation.They tried multiple times but the results remain the same.So, they removed the device back.The deformed device was never released from the delivery cable while inside the patient.The procedure got aborted.There was no report of any interaction with cardiac structures during deployment and no report of any angulation/kink noticed with the delivery system.The patient remained hemodynamically stable throughout the procedure.There was no clinically significant delay in procedure and no adverse patient effects.The patient was reported stable.
|
| |
|
Search Alerts/Recalls
|