| Model Number |
P1170G |
| Medical Device Problem Code |
Unintended Movement (3026)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/10/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The baxter technician found the caregiver membrane needed to be replaced.Per the baxter service manual the bed should be subject to an effective maintenance program.An annual service of the bed is advised in order to maintain its characteristics and performance.Make sure the head and foot siderails are not bent or twisted.Make sure the latch mechanism operates correctly.An audible click must be heard.Remove the siderail cover, and make sure the mounting screws are tight.Repair or replace the siderails as necessary.A search of the baxter maintenance records showed baxter performed preventative maintenance on this bed in apr 22, 2024.It is unknown if the facility performed any other preventative maintenance on this bed.The technician replaced the caregiver membrane to resolve the reported event.Based on this information, no further action is required.
|
| |
|
Event or Problem Description
|
|
On 10-feb-2026, a customer contacted technical service to report that careassist es (product code p1170g0000313, serial number (b)(6)), had head moving up by itself.There was no patient/user injury, medical intervention, symptom, or adverse event reported.
|
| |
|