| Model Number |
2120 |
| Medical Device Problem Code |
Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/29/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
One device was received for evaluation.Visual inspection was performed and noted a torn downstream occlusion seal.The seal was replaced.The device then passed all functional tests.The service history review had no indication that the complaint was related to a service of the device within the review period.
|
| |
|
Event or Problem Description
|
|
The device evaluation noted a torn downstream occlusion seal.The event occurred during testing.There was no patient involvement and harm.
|
| |
|