• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL AGILIS¿ NXT STEERABLE INTRODUCER DUAL-REACH¿ 61 CM SHEATH, MEDIUM CURL, 8.5F; INTRODUCER, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL AGILIS¿ NXT STEERABLE INTRODUCER DUAL-REACH¿ 61 CM SHEATH, MEDIUM CURL, 8.5F; INTRODUCER, CATHETER Back to Search Results
Model Number G408319
Medical Device Problem Code Gas/Air Leak (2946)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/13/2026
Type of Reportable Event Malfunction
Event or Problem Description
During an af procedure, a feeling like air being drawn in was noted when aspirating the sheath.The sheath was replaced and the procedure was completed with no adverse consequences to the patient.
 
Additional Manufacturer Narrative
Additional information: d9, g3, h2, h3, h6, h11.One 8.5f agilis steerable introducer sheath was received for evaluation.An event of a leak was reported.Functional testing determined a leak was noted in the hemostasis valve.The cap was removed from the hemostasis hub, and the hemostasis seals were microscopically inspected.Tearing, resulting in a hole, was noted in the proximal and distal seals.The batch record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the damaged seals and subsequent leak remains unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AGILIS¿ NXT STEERABLE INTRODUCER DUAL-REACH¿ 61 CM SHEATH, MEDIUM CURL, 8.5F
Common Device Name
INTRODUCER, CATHETER
Manufacturer (Section D)
ABBOTT MEDICAL
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS  1897-4050
Manufacturer (Section G)
ABBOTT MEDICAL
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS   1897-4050
Manufacturer Contact
wendy mccarty
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key24532578
Report Number3008452825-2026-00098
Device Sequence Number15677313
Product Code DYB
UDI-Device Identifier05414734206129
UDI-Public(01)05414734206129(10)11082136(17)280930
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K061363
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/24/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberG408319
Device Catalogue NumberG408319
Device Lot Number11082136
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/28/2026
Supplement Date Received by Manufacturer03/17/2026
Initial Report FDA Received Date03/06/2026
Supplement Report FDA Received Date03/24/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/09/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-