• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN AG JUAREZ (MMJ) TAPERGUARD; TUBE, TRACHEAL (W/WO CONNECTOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN AG JUAREZ (MMJ) TAPERGUARD; TUBE, TRACHEAL (W/WO CONNECTOR) Back to Search Results
Model Number 18765S
Medical Device Problem Code Use of Device Problem (1670)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/27/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
It was reported that it was difficult to inject air into the endotracheal tube's cuff prior to use.There was no patient involved.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TAPERGUARD
Common Device Name
TUBE, TRACHEAL (W/WO CONNECTOR)
Manufacturer (Section D)
COVIDIEN AG JUAREZ (MMJ)
1181 hennequen ave.
juarez,chi 32575
MX  32575
Manufacturer (Section G)
COVIDIEN AG JUAREZ (MMJ)
1181 hennequen ave.
juarez,chi 32575
MX   32575
Manufacturer Contact
mike bocchino
8200 coral sea st ne
mounds view, MN 55112-4391
8006464633
MDR Report Key24532939
Report Number9681384-2026-00276
Device Sequence Number21066298
Product Code BTR
UDI-Device Identifier10884521620179
UDI-Public(01)10884521620179(17)300820(10)25H0999JZX
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K090352
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/24/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number18765S
Device Catalogue Number18765S
Device Lot Number25H0999JZX
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/05/2026
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/02/2026
Supplement Date Received by Manufacturer06/03/2026
Initial Report FDA Received Date03/06/2026
Supplement Report FDA Received Date06/24/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/20/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-