| Catalog Number |
SSF-415 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
02/19/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported ¿mri was interpreted as showing a ruptured implant.When removed the implant, it was found to be intact¿.This record is for an unknown side.Device was explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: per the analysis performed, the assessments of the complaints and any potential manufacturing issue are displayed along with any further actions required: ¿ rupture: not observed through visual inspection.None of the other observations performed during the device analysis creases, wear abrasion and deformation are found to be potentially related to the manufacturing process, and, therefore, no further actions are required for these observations.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported ¿mri was interpreted as showing a ruptured implant.When removed the implant, it was found to be intact¿.This record is for an unknown side.Device was explanted.
|
| |
|